Skip to Content To homepage

Versa 20/Hitachi

Versatile Peltier-driven temperature control for Hitachi Spectrophotometers
Product Overview

The Versa 20/Hitachi is a versatile design, providing temperature control to a single cuvette in the forward sample beam of the spectrophotometer. The Versa 20/Hitachi is ideal for work in the biological range of 0 °C to 110 °C, either to observe absorbance spectra at precisely maintained temperatures, or for temperature ramping, to observe how absorbance changes with temperature. To work at cold temperatures, purchase a Windowed Jacket, purge with dry gas and work down to -40 °C. For even more demanding work, purchase the extended temperature option and run the Versa 20/Hitachi up to 150 °C or down to below -40 °C. The Versa 20/Hitachi replaces the sample tray (base and front panel) of the spectrophotometer, with utilities brought conveniently to the front panel. This unit may be controlled using the menu button on the front of the TC 1/Single controller or through program T-App available from Quantum Northwest.

Product Features

Rapid, precise temperature control
Optional extended temperature range
Windowed jacket available for temperature extremes
Variable speed magnetic stirring
Dry gas purge for work at low temperature
Complete sample tray for mounting
Probe input for sample temperature
NIST-traceable temperature calibration

  • Technical Specifications
    Number of cuvettes controlled
    1
    Typical use
    Absorbance
    Standard cuvette size (outside)
    12.5 mm x 12.5 mm
    Optical ports per cuvette
    2
    Optical port dimensions
    10 mm high x 10 mm wide
    Cuvette z-height
    8.5 or 15 mm for most models
    Windowed jacket available
    2 windows
    Normal factory set temperature range
    -40 °C to +105 °C
    Temperature range easily achievable
    -15 °C to +105 °C
    Range with special techniques
    -40 °C to +105 °C
    Extended factory set temperature range
    -55 °C to +150 °C
    Extended temperature range achievable
    -50 °C to +150 °C
    Temperature precision
    ±0.02 °C
    Temperature accuracy
    Better than ±0.15 °C from 0 ⁰C to +80 ⁰C
    Accuracy greater range
    Better than ±0.3 ⁰C from ‐20 ⁰C to +110 ⁰C
    Temperature reproducibility
    Better than ±0.07 °C
    Thermistor probes accepted
    Series 400 or Series 500
    Magnetic stirring motor type
    Stepper
    Magnetic stirring speed range
    1-2500 rpm
    Default stirring speed
    1200 rpm
    Stirring speed best performance
    60-1800 rpm
  • Basic Package

    Versa 20 – Temperature-controlled cuvette holder (standard z-height = 15 mm)
    TC 1/Single – Temperature Controller
    Sample compartment tray with front panel utility access
    Reference cuvette holder without temperature control
    BATH 10 – Circulator
    NIST-traceable calibration and performance data
    Accessories – Magnetic stir bar, tubing, cables and opaque cap

  • Accessories
Other Products to Consider