Skip to Content To homepage

t2x2 Sport/Cary5000

Dual temperature control for the Cary 4000, 5000, 6000i and 7000 UV-Vis-NIR Spectrophotometers
Product Overview

The t2x2 Sport/Cary5000 is a compact pair of cuvette holders providing Peltier temperature control for both the sample and reference cuvettes. Use it for temperature ranges of about -15 °C to 110 °C. It is ideal for most biological studies, either to observe absorbance spectra at precisely maintained temperatures, or for temperature ramping, to observe how absorbance changes with temperature. If necessary, run the sample and reference cuvettes at different temperatures. The mounted t2x2 Sport/Cary5000 replaces both the sample and reference cuvette holders normally provided with the spectrophotometer. Two options are available for bringing utilities (tubes and wires) out of the sample compartment. For laboratories that have the Agilent extended “nose” assembly, we can provide a simple access plate that mounts below the nose. Otherwise, you may purchase our PANEL/5000/x2 option, which permits connections of the front of the spectrophotometer. Operate the TC 1/Dual Temperature Controller using the menu button or control via program T-App. Use an ADL program for temperature ramping with WinUV ADL Shell software.

Product Features

Temperature control for both sample and reference
Variable speed magnetic stirring
Dry gas purge for work at low temperature
Bases and posts for mounting in the spectrophotometer
Cuvette lifters for removal of cuvettes
Probe input for sample temperature
NIST-traceable temperature calibration
ADL program for temperature ramping

  • Technical Specifications
    Number of cuvettes controlled
    2
    Typical use
    Absorbance (sample and reference)
    Standard cuvette size (outside)
    12.5 mm x 12.5 mm
    Optical ports per cuvette
    2
    Optical port dimensions
    12 mm high x 10 mm wide
    Windowed jacket available
    no
    Factory set temperature range
    -40 °C to +110 °C
    Temperature range easily achievable
    -15 °C to +110 °C
    Temperature precision
    ±0.02 °C
    Temperature accuracy
    Better than ±0.15 °C from 0 ⁰C to +80 ⁰C
    Accuracy full range
    Better than ±0.3 ⁰C from ‐15 ⁰C to +110 ⁰C
    Temperature reproducibility
    Better than ±0.07 °C
    Magnetic stirring motor type
    Stepper under each cuvette
    Magnetic stirring speed range
    1-2500 rpm
    Default stirring speed
    1200 rpm
    Stirring speed best performance
    60-1800 rpm
  • Basic Package

    t2x2 Sport – Pair of cuvette holders
    TC 1/Dual – Temperature Controller
    Base plates, mounting posts (standard z-height = 15 mm), utility access panel
    BATH 10 – Circulator
    NIST-traceable calibration and performance data
    ADL program for temperature ramping
    Accessories – Two magnetic stir bars, tubing and cables

  • Accessories
Other Products to Consider
Back to top