Skip to Content To homepage

Luma 40/Cary5000

Temperature-controlled cuvette holder with four optical ports for the Cary 4000, 5000, 6000i and 7000 UV/Vis/NIR Spectrophotometers
Product Overview

The Luma 40/Cary 5000 has four optical ports around the cuvette. Use it for absorbance measurements when your experiment requires access at 90 degrees for photochemical activation. The Luma 40/Cary5000 is a versatile design that is used in the Cary 4000, 5000 and 6000i UV/Vis/NIR Spectrophotometers. Purchase the extended temperature option and run the Luma 40/Cary5000 up to 150 °C for non-aqueous samples. Add a Windowed Jacket, purge with dry gas and run the temperature down to at least -40 °C. The Luma 40 is also ideal for work in the biological range of 0 °C to 110 °C. Use it to observe absorbance spectra at precisely maintained temperatures, or for temperature ramping, to observe how absorbance changes with temperature. Two options are available for bringing utilities (tubes and wires) out of the sample compartment. For laboratories that have the Agilent extended “nose” assembly, we can provide a simple access plate that mounts below the nose. Otherwise, you may purchase our PANEL/5000 option, which permits connections of the front of the spectrophotometer. Operate the TC 1/Single Temperature Controller using the menu button or control via program T-App. Use an ADL program for temperature ramping with WinUV ADL Shell software.

Product Features

Four optical ports
Rapid, precise temperature control
Optional extended temperature range
Windowed jacket available for temperature extremes
Variable speed magnetic stirring
Dry gas purge for work at low temperature
Base and post for mounting in the spectrophotometer
Mounting base and light-tight front panel
Probe input for sample temperature
NIST-traceable temperature calibration
ADL program for temperature ramping

  • Technical Specifications
    Number of cuvettes controlled
    1
    Typical use
    Fluorescence
    Standard cuvette size (outside)
    12.5 mm x 12.5 mm
    Optical ports per cuvette
    4
    Optical port dimensions
    12 mm high x 10 mm wide
    Built-in optical slit holders
    4 per cuvette
    Cuvette z-height
    8.5 or 15 mm for most models
    Windowed jacket available
    2, 3 or 4 windows
    Normal factory set temperature range
    -40 °C to +105 °C
    Temperature range easily achievable
    -15 °C to +105 °C
    Range with special techniques
    -40 °C to +105 °C
    Extended factory set temperature range
    -55 °C to +150 °C
    Extended temperature range achievable
    -50 °C to +150 °C
    Accuracy greater range
    Better than ±0.3 ⁰C from ‐20 ⁰C to +110 ⁰C
    Temperature reproducibility
    Better than ±0.07 °C
    Thermistor probes accepted
    Series 400 or Series 500
    Magnetic stirring motor type
    Stepper
    Magnetic stirring speed range
    1-2500 rpm
    Default stirring speed
    1200 rpm
    Stirring speed best performance
    60-1800 rpm
  • Basic Package

    Luma 40 – Temperature-controlled cuvette holder (standard z-height = 15 mm)
    Base plate and utility access panel
    TC 1/Single – Temperature Controller
    Bath 10 – Circulator
    ADL program for temperature ramping
    NIST-traceable calibration and performance data
    Accessories – Magnetic stir bar, tubing, cables, optical slits and opaque cap

  • Accessories
Other Products to Consider
Back to top