LC
600™
The LC 600™ is a temperature-controlled six-position linear
changer for use in absorbance spectrophotometers (including circular
dichroism). Magnetic stirring is provided under each cuvette and
a dry gas purge may be used to dry gas on the cuvette windows to
prevent condensation when working at low temperatures.
| • |
Rapid,
precise Peltier control over a wide range
of temperatures* |
| • |
Variable
speed magnetic stirring under each cuvette |
| • |
Dry
gas purge |
| • |
Optional
remote computer control |
| • |
TC
125 temperature-controller included |
Description
The insulated temperature-controlled body of the LC 600™ holds
up to six standard 10 x 10 mm square cuvettes. Two 10 x 10 mm optical
ports on opposite sides of each cuvette permit the passage of the
spectrophotometer beam. The height of the optical center from the
bottom of the cuvette is 8.5 mm to permit the use of standard microcuvettes
requiring this dimension. A stirring magnet below each cuvette
permits magnetic stirring to minimize temperature gradients in
the cuvette.
The temperature-regulated cuvette holder contains a thermoelectric
cooler (Peltier device) for controlling the temperature. Circulating
water (or other fluid) passes through a heat exchanger to draw off
the heat generated by the cooler. Dry gas may be directed through
small holes in the optical ports next to the exposed cuvette windows.
The dry gas may be used to limit condensation when working at low
temperatures. An opaque lid can be placed over the tower and cuvette.
The cuvette holder is matched to a specific temperature controller
and calibrated against a NIST-traceable thermometer. Calibration
data are provided.
Translation to change samples uses a lead screw driven by a stepping
motor.The LC 600™ requires external computer control to translate.
This may be accomplished through the spectrometer software (if
such control is currently available) or through the Serial Control
option.
This linear changer is currently available for Olis instruments
and the Cary 300. If there is adequate space, we will be pleased
to adapt
it to any other spectrometer.

Sample
Holder Temperature Ranges
|